Frequently Asked Questions

General chemical fume hood FAQ

When should I use a fume hood?

Chemical fume hoods safely vent toxic fumes and highly corrosive chemicals away from your lab environment. You need a hood if your work involves using any of the following.

  • Chemicals with a National Fire Protection Association (NFPA) health rating of 3 or 4
  • Toxic volatile materials such as chloroform or formaldehyde
  • Flammable chemicals
  • Carcinogens or particularly hazardous substances
  • Procedures that may create an aerosol of a toxic substance
  • Reactive or explosive materials or chemicals that may spatter
  • Toxic gases, including NH3, CO, F2, Cl2, H2S and NO2
  • Odorous materials, whether hazardous and non-hazardous
When should I use a polypropylene fume hood?

The primary purpose of a polypropylene hood is to avoid metal deterioration in highly corrosive environments. A standard hood may be used in the same applications, but the key advantage of a polypropylene hood is its resistance to corrosion.

What can I store in the hood?

Hoods are designed for procedural work and not storage of any kind within the work zone. Storing equipment, chemicals or other large objects may disrupt airflow and eliminate the hood’s intended functionality.

What if I need to store large pieces of equipment in the hood?

If you must store large equipment in the hood that can potentially block the rear baffle, raise the equipment off the surface of the hood. This allows for proper airflow to reach the back of the hood.

Why should I keep work away from the front of the hood?

When chemical vapors are near the front of the hood, they aren’t efficiently captured and can escape into the room. To maintain proper containment, chemicals should be kept at least 6 inches back from the front of the work zone.

What is face velocity?

Face velocity is the speed at which room air is drawn into the hood through the opening, or “face.”  The face of the hood is the opening where air capture takes place. Face velocity is measured during performance testing at the factory and after installation and indicates the hood’s performance.

What factors affect vapor capture in a fume hood?

As you work at the hood, air is drawn in a laminar flow through the face of the hood. Air velocity is one of the factors determining how well vapors and particles are captured and exhausted from the hood work zone. Vapor capture is also affected by airfoils, hood shape, configuration, equipment inside the hood, and cross drafts from foot traffic, proximity to doors and room vents.

Why is 100 fpm used as the ideal face velocity?

An optimal face velocity falls between 80 and 120 linear feet per minute (fpm). At lower velocities, the hood may not capture and remove vapors at a sufficient rate. At higher velocities, airflow can become turbulent, producing eddies and backflows that hinder the rate of removal or cause vapors to escape. Often, the best measurement of hood effectiveness is through visualization with dry ice or other visible fumes released in the hood.

Is using flammable chemicals in a fume hood with lights inside safe?

Yes. The lights inside hoods are sealed to prevent flammable vapors and are considered “intrinsically safe.” This means all electrical components in new hoods are designed so that flammable vapors should not be a fire hazard.

However, if you use other electrical equipment or heating apparatus in the hood, they may pose a fire risk. Always ensure they are intrinsically safe, or use them outside the hood. Take care with all heating devices and flammable solvents. Avoid open flames, and carefully monitor the temperature of other heating elements.

What should I consider when selecting a standalone or “ductless” fume hood that doesn’t vent to the outside?

Filters used in ductless hoods are an important consideration. Hoods that do not vent outside the building rely on filters inside the cabinet. These filters remove vapors and toxins from the air inside the hood before discharging into the lab air. Filters do not completely capture all vapors, and some vapors can end up in the lab’s breathing air. Also, contaminants that build up on the filters can become a hazard. To maintain safety, it’s essential to establish a maintenance schedule to replace and dispose of filters, usually managed by a contractor. 

Technical and compliance FAQ

What’s the difference between the exhaust requirements CFM and the certification value CFM listed?

The exhaust requirement CFM represents the airflow rate used in calculations to balance the room’s ventilation where the hood will be installed. This number is 1.1 times the certification value. The certification value CFM is the airflow rate at which the hood is designed and tested.

Airflow characteristics are for a 10-inch opening. Could this data be revised to reflect an 18-inch opening?

The hood is designed to operate with a 10-inch opening to provide the necessary protection for both product and personnel (as required by the National Safety Foundation Standard No. 49). We do not provide numbers for an 18-inch opening or certify that the hood would perform safely with a larger opening.  

Casework FAQ

Do you offer a polypropylene flammable base cabinet?

We don’t. National Fire Protection Association (NFPA) 30 code requires constructing flammable liquid storage cabinets so that, during a 10-minute fire test, the internal temperature at the center and one inch from the top does not exceed 325 degrees Fahrenheit.

Is all polypropylene flame-retardant?

Not all polypropylene is flame-retardant, but we offer several polypropylene options that meet non-metallic flammability requirements. 

Our standard polypropylene carries a UL 94 HB rating from Underwriters Laboratories. A horizontal burn (HB) flame rating indicates that the material was tested in a horizontal position and found to burn at a rate less than a specified maximum. 

We also offer polypropylene with a UL-94 V-O rating, meaning it was tested in a vertical position and self-extinguished within a specified time after the ignition source was removed. The rating also indicates whether the test specimen dripped flaming particles that ignited a cotton indicator located below the sample. 

The highest-rated material we offer carries an FM4910 certification, which is designed for cleanroom occupancies. FM4910 evaluates the materials on their fire propagation behavior and potential for smoke contamination using two key indices: Fire Propagation Indes (FPI) and Smoke Damage Index (SDI).

Can locks be added to cabinets?

Yes. We offer several lock options, including epoxy-coated, non-ferrous and magnetic resonance imaging (MRI)-rated material.

Do you offer Revit files?

No, but we can provide DXF 2-D files compatible with most CAD files. 

Do you provide installation?

No, but we work with several laboratory equipment providers who may offer installation. More information is available on our Architects & Lab Planning page or contact us.  

Can you accommodate ADA requirements?

Yes. We can modify most units to comply with ADA requirements or custom requirements.

Have a question not listed?

Get in touch with us
LabAire Systems

LabAire Systems
9560 Newton Ave S
Bloomington, MN 55431

Phone: (877) 410-4910

Careers

  • LinkedIn
Contact Us

Products

  • Polypropylene Fume Hoods
  • Polypropylene Casework
  • Polypropylene Wet Process Benches
  • Lab Accessories

RESOURCEs

  • Guide specifications
  • Industry resources
  • Order sample material
  • Catalogs and product data sheets

ISO 9001:2015 Certified

Diversified American Manufacturing

Copyright © 2026 LabAire Systems. A division of Activar Plastic Products Group, Inc. All Rights Reserved.

Terms | Privacy Policy